CHAPTER II: ILLINOIS EMERGENCY MANAGEMENT AGENCY SUBCHAPTER b: RADIATION PROTECTION PART 370 QUALITY STANDARDS AND CERTIFICATION REQUIREMENTS FOR FACILITIES PERFORMING MAMMOGRAPHY SECTION 370.TABLE A MAMMOGRAPHY DOSE EVALUATION TABLE
Section 370.TABLE A Mammography Dose Evaluation Table
This Table is used to determine the mean glandular dose in milligrays delivered by 25.8 mC/kg (or millirad) delivered by 1 R in air incident on a 4.2 centimeter thickness compressed breast of average density (50 percent adipose and 50 percent glandular tissue). Values listed are for the first half-value layer (HVL) in millimeters of aluminum (mm Al), for x-ray tube target-filter combinations of molybdenum/molybdenum (Mo/Mo) and tungsten/ aluminum (W/Al). Linear extrapolation or interpolation shall be made for any HVL not listed.
Mean Glandular Dose in milligrays for 25.8 mC/kg (or millirad for 1 R) Entrance Exposure for a 4.2 Centimeter Compressed Breast of Average Density
AGENCY NOTE: Adapted from: Mammography Quality Control Manual: Medical Physicist's Section, Revised Edition, 1999.
(Source: Amended at 33 Ill. Reg. 2224, effective January 23, 2009) |